FDA recall record

Siemens Medical Solutions USA Ysio X.pree recall — FDA Z-1346-2021

A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector…

Class IIclassification
Z-1346-2021official record ID
Apr 14, 2021report date

Recall record at a glance

Official recordZ-1346-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateApr 14, 2021
Recall initiation dateMar 10, 2021
Quantity12 units
LocationMalvern, PA, United States

Official product description

Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body including the skull, chest, abdomen, and extremities.

Why the product was recalled

A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted

Distribution information

US Nationwide distribution in the states of FL, MN, OH, PA, TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.