FDA recall record

Angiodynamics XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat recall — FDA Z-0505-2020

Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock…

Class IIclassification
Z-0505-2020official record ID
Nov 27, 2019report date

Recall record at a glance

Official recordZ-0505-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAngiodynamics Inc. (Navilyst Medical Inc.)
Report dateNov 27, 2019
Recall initiation dateSep 18, 2019
Quantity582
LocationGlens Falls, NY, United States

Official product description

XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

Why the product was recalled

Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in a delay to the implantation procedure.

Distribution information

Worldwide distribution - US Nationwide in the state of CA, CT, FL, GA, IA, IN, KS, KY, MI, MO, MS, MT, NE, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI and country of Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.