Fenwal VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a recall — FDA Z-2064-2019

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC…

Class Iclassification
Z-2064-2019official record ID
Aug 21, 2019report date

Recall record at a glance

Official recordZ-2064-2019
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmFenwal Inc.
Report dateAug 21, 2019
Recall initiation dateJun 24, 2019
Quantity9461 units
LocationLake Zurich, IL, United States

Official product description

VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.

Why the product was recalled

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.