FDA recall record

Hill-Rom Volker 2080, 2082, S 280 and S 282 series beds with lateral release button recall — FDA Z-2376-2016

The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly…

Class IIclassification
Z-2376-2016official record ID
Aug 17, 2016report date

Recall record at a glance

Official recordZ-2376-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHill-Rom, Inc.
Report dateAug 17, 2016
Recall initiation dateJun 14, 2016
Quantity24262
LocationBatesville, IN, United States

Official product description

Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.

Why the product was recalled

The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury.

Distribution information

Worldwide Distribution - US Nationwide and the countries of: Canada, AUSTRALIA, CHINA, HONG KONG, THAILAND, POLAND, RUSSIAN FEDERATION, SLOVENIA, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, ICELAND, IRELAND, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, SPAIN, SWITZERLAND, and UNITED KINGDOM.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 6, 2021

Hill-Rom LikoStretch Mod 600 IC recall — FDA Z-2601-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.
Medical DeviceOct 6, 2021

Hill-Rom LikoStretch 1900 recall — FDA Z-2600-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.