FDA recall record

Maquet SAS VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP recall — FDA Z-2502-2018

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that…

Class IIclassification
Z-2502-2018official record ID
Jul 25, 2018report date

Recall record at a glance

Official recordZ-2502-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMaquet SAS
Report dateJul 25, 2018
Recall initiation dateJun 14, 2018
Quantity151
LocationOrleans, France

Official product description

VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.

Why the product was recalled

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

Distribution information

Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, CO, CT, FL, GA, IA, ID, KS, KY, LA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, and WA. Foreign distribution to Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Tanzania, Yemen, and Zimbabwe.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.