FDA recall record

Volcano s5i/s5ix/s5iz/CORE Intravascular Ultrasound Imaging and Pressure System recall — FDA Z-2044-2015

A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This…

Class IIclassification
Z-2044-2015official record ID
Jul 15, 2015report date

Recall record at a glance

Official recordZ-2044-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVolcano Corp
Report dateJul 15, 2015
Recall initiation dateSep 23, 2013
Quantity227 total units - all models
LocationSan Diego, CA, United States

Official product description

Volcano s5i/s5ix/s5iz/CORE Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5i -806300-020 ; s5ix/s5iz - 807400-001; CORE - 400-0 I 00.02. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.

Why the product was recalled

A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.

Distribution information

Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceApr 20, 2022

Volcano MM-TSM recall — FDA Z-0917-2022

Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore,…

Class IIIFDA: OngoingVolcano Corp