FDA recall record

Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 recall — FDA Z-2131-2015

A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

Class IIclassification
Z-2131-2015official record ID
Jul 29, 2015report date

Recall record at a glance

Official recordZ-2131-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVolcano Corporation
Report dateJul 29, 2015
Recall initiation dateJun 22, 2015
QuantityTotal of 4007 devices, all models
LocationRancho Cordova, CA, United States

Official product description

Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5. Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature

Why the product was recalled

A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

Distribution information

Worldwide Distribution-US (nationwide) and regions in Europe and the countries of Japan, Canada, South Africa, Taiwan, Thailand, Singapore, Philippines, Australia, New Zealand, El Salvador, Malaysia, Mexico, Lebanon, Kuwait, South Korea, Kenya, Jordan, Iran, India, Israel, Hong Kong, Egypt, Algeria, Dominican Republic, Costa Rica, Chile, Brazil, Bolivia, and Argentina

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.