FDA recall record

LivaNova USA VNS Therapy, SenTiva, Model # 1000, (01)05425025750405, Rx Only for vagal nerve recall — FDA Z-0561-2020

Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to…

Class Iclassification
Z-0561-2020official record ID
Jan 1, 2020report date

Recall record at a glance

Official recordZ-0561-2020
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmLivaNova USA Inc.
Report dateJan 1, 2020
Recall initiation dateAug 22, 2019
Quantity1687
LocationHouston, TX, United States

Official product description

VNS Therapy, SenTiva, Model # 1000, (01)05425025750405, Rx Only for vagal nerve stimulation (VNS) therapy

Why the product was recalled

Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.

Distribution information

U.S.: MN, IL, AK, NY, FL, PA, AR, CA, NC, WI, AZ, KY, TX, MA, MT, NE, IA, WV, CO, AL, MI, LA, VA, OH, MO, IN, NJ, GA, TN, MS, DC, SC, CT, UT, OK, MD, ID, KS, DE, OR, ND, WA, NM, PR. OUS (foreign): AT, CH, DE, ES, FI, GB, IE, NL, NO, PT, SE, IT

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.