LivaNova USA VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 recall — FDA Z-1563-2020

False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3)…

Class IIclassification
Z-1563-2020official record ID
Apr 1, 2020report date

Recall record at a glance

Official recordZ-1563-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmLivaNova USA Inc.
Report dateApr 1, 2020
Recall initiation dateFeb 4, 2020
Quantity2683
LocationHouston, TX, United States

Official product description

VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.

Why the product was recalled

False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.

Distribution information

U.S.: LA, ID, AZ, UT, KY, VA, TN, NY, SC, IL, IN, NC, WA, CA, OH, FL, PA, NJ, CO, TX, RI, MD, HI, OR, MA, NE, MO, OK, MS, MN, SD, VT, AR, WV, WI, MI, AL, KS, DE, NM, GA, ND, AK, MT, ME, NV, DC, MO, WY, IA, CT. Canada, and Israel.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 12, 2025

LivaNova USA SenTiva, REF: 1000 recall — FDA Z-1072-2025

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in…

Class IIFDA: OngoingLivaNova USA, Inc.