FDA recall record

Ortho-Clinical Diagnostics VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For recall — FDA Z-0878-2023

Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results

Class IIclassification
Z-0878-2023official record ID
Jan 11, 2023report date

Recall record at a glance

Official recordZ-0878-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOrtho-Clinical Diagnostics, Inc.
Report dateJan 11, 2023
Recall initiation dateNov 17, 2022
Quantity4,158 units
LocationRochester, NY, United States

Official product description

VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) Catalog Number: 6802892

Why the product was recalled

Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results

Distribution information

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Bermuda HM12, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.