FDA recall record

Ortho-Clinical Diagnostics VITROS Chemistry Products VersaTips, Catalog No. 6801715 A recall — FDA Z-0735-2022

The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected…

Class IIclassification
Z-0735-2022official record ID
Mar 16, 2022report date

Recall record at a glance

Official recordZ-0735-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOrtho-Clinical Diagnostics, Inc.
Report dateMar 16, 2022
Recall initiation dateDec 10, 2021
Quantity4125 units
LocationRochester, NY, United States

Official product description

VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.

Why the product was recalled

The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.

Distribution information

Domestic distribution nationwide. Foreign distribution to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.