FDA recall record

Ortho-Clinical Diagnostics VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use recall — FDA Z-0510-2024

Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may…

Class IIclassification
Z-0510-2024official record ID
Dec 13, 2023report date

Recall record at a glance

Official recordZ-0510-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOrtho-Clinical Diagnostics, Inc.
Report dateDec 13, 2023
Recall initiation dateOct 25, 2023
Quantity622 units 300 Slides (5 cartridge containing 60 slides)
LocationRochester, NY, United States

Official product description

VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively measure total carbon dioxide (CO2) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600 XT 3400 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Product Code: 8262396

Why the product was recalled

Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.

Distribution information

Nationwide Foreign: Bermuda Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.