FDA recall record

Ortho-Clinical Diagnostics VITROS Chemistry Products CRP Slides- In vitro diagnostic recall — FDA Z-0535-2025

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L…

Class IIclassification
Z-0535-2025official record ID
Dec 4, 2024report date

Recall record at a glance

Official recordZ-0535-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOrtho-Clinical Diagnostics, Inc.
Report dateDec 4, 2024
Recall initiation dateOct 11, 2024
Quantity7717 units
LocationRochester, NY, United States

Official product description

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740

Why the product was recalled

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

Distribution information

Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: not stated.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.