FDA recall record

Ortho-Clinical Diagnostics VITROS Chemistry Product Performance Verifier II, Product Code recall — FDA Z-2367-2020

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed…

Class IIclassification
Z-2367-2020official record ID
Jun 24, 2020report date

Recall record at a glance

Official recordZ-2367-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOrtho-Clinical Diagnostics
Report dateJun 24, 2020
Recall initiation dateMay 11, 2020
LocationRochester, NY, United States

Official product description

VITROS Chemistry Product Performance Verifier II, Product Code 8067324, UDI # 10758750004317 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Why the product was recalled

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Distribution information

Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.