FDA recall record

Organon Teknika Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID recall — FDA Z-0962-2019

This notice has been initiated due to potential for compromised test card Top Seal Integrity which could: - yield false resistance for antibiotics on the…

Class IIclassification
Z-0962-2019official record ID
Mar 6, 2019report date

Recall record at a glance

Official recordZ-0962-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOrganon Teknika Inc.
Report dateMar 6, 2019
Recall initiation dateJan 11, 2019
Quantity129,321 cartons
LocationDurham, NC, United States

Official product description

Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID, REF 21347, d) AST-GNOR, REF 22000, e) AST-GP67, REF 22226, f) AST-P612, REF 22359, g) AST-XN01, REF 410025, h) AST-N222, REF 413083, i) AST-N223, REF 413110, j) AST-N226, REF 413143, k) AST-N248, REF 413397, l) ASR-GN67, REF 413399, m) AST-GN69, REF 413400, n) AST-GN73, REF 413404, o) AST-GN79, REF 413436, p) AST-GN82, REF 413439, q) AST-N261, REF 413753, r) AST-XN06, REF 413944, s) AST-N275, REF 414313, t) AST-GP75, REF 415670, u) AST-N300, REF 416241, v) AST-N308, REF 413913, w) AST-N314, REF 417027, x) AST-N318, REF 417952, y) AST-P643, REF 418671, z) AST-YS08, REF 420739, aa) AST-N348, REF 420856, bb) AST-ST02, REF 420915, cc) AST-ST03, REF 421040, dd) AST-N355, REF 421351, ee) AST-P654, REF 421912, ff) AST-P655, REF 421913, gg) AST-N371, REF 422024, hh) AST-N372, REF 422241, ii) AST-N379, REF 423052,

Why the product was recalled

This notice has been initiated due to potential for compromised test card Top Seal Integrity which could: - yield false resistance for antibiotics on the AST panel - cause a false negative ESBL Test - result in a false positive urea (URE) reaction on ID cards

Distribution information

Worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.