FDA recall record

Zimmer Biomet Spine Vitality Spinal Fixation System recall — FDA Z-0866-2019

The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing…

Class IIclassification
Z-0866-2019official record ID
Mar 6, 2019report date

Recall record at a glance

Official recordZ-0866-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZimmer Biomet Spine Inc.
Report dateMar 6, 2019
Recall initiation dateJul 30, 2018
Quantity16,429 devices (Thoracolumbosacral pedicle screw system,Spinal Interlaminal & Spinal Intervertebral Body, both appliances for fixation)
LocationWestminster, CO, United States

Official product description

Vitality Spinal Fixation System. Thoracolumbosacral pedicle screw system (non-sterile). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. For use in Orthopedic medical specialty area. Product Usage: The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue. These devices are indicated as an adjunct to fusion for various indications

Why the product was recalled

The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications

Distribution information

Distributed Nationwide: AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, LA, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA and WI.**** International distribution: Australia and EMEA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.