FDA recall record

Galil Medical Visual-ICE Cryoablation System, FPRCH6000 recall — FDA Z-0404-2020

Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing…

Class IIclassification
Z-0404-2020official record ID
Nov 20, 2019report date

Recall record at a glance

Official recordZ-0404-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGalil Medical, Inc.
Report dateNov 20, 2019
Recall initiation dateOct 25, 2019
Quantity49 systems
LocationSaint Paul, MN, United States

Official product description

Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively , by the use of Galil Medical I-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing.

Why the product was recalled

Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.

Distribution information

AK, AL, AR, AZ, CA, FL, GA, HI, KY, MA, MI, MN, NC, ND, NY, OH, PA, SC, TN, UT, VA, Puerto Rico OUS to include: Argentina, Canada, Taiwan, United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.