FDA recall record

Lacrimedics VisiPlugST for the Lacrimal Efficiency Test recall — FDA Z-1081-2015

Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf…

Class IIIclassification
Z-1081-2015official record ID
Feb 18, 2015report date

Recall record at a glance

Official recordZ-1081-2015
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmLacrimedics Inc.
Report dateFeb 18, 2015
Recall initiation dateNov 26, 2014
Quantity781 units
LocationDuPont, WA, United States

Official product description

VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 1815, 1903, 1904, and 1905. This device is cylindrical in shape, approximately 1.75-2.00mm in length, and available in three sizes: 0.3mm, 0.4mm and 0.5mm. Plugs are place in the upper and/or lower horizontal canaliculus. Two plugs are placed into a foam carrier and nitrogen sealed inside a foil pouch. The foil pouch is sent to sterilizer for gamma sterilization. Two sterile pouches are packaged in a box with an IFU.

Why the product was recalled

Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.

Distribution information

Worldwide Distribution-US (nationwide) including the states of AR, CA, FL, ID, IL, KS, MA, MI, NC, NJ, NY, OR, PA, and TX and to the countries of Australia, Brazil, Czech Republic, France, Ireland, Mexico, Peru, Switzerland, and United Kingdom.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.