FDA recall record

Sequent Medical VIA 27 Microcatheter, Part No recall — FDA Z-0114-2017

Sequent Medical is recalling the VIA 27 Microcatheter because it may not adequately re-sheath the interventional device into the catheter as designed.

Class IIclassification
Z-0114-2017official record ID
Nov 2, 2016report date

Recall record at a glance

Official recordZ-0114-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSequent Medical Inc.
Report dateNov 2, 2016
Recall initiation dateSep 23, 2016
Quantity618 units
LocationAliso Viejo, CA, United States

Official product description

VIA 27 Microcatheter, Part No. FG27-154-01, Model No. VIA-27-154-01 The VIA 27 is intended for the introduction of non-liquid interventional devices (such as stents/flow diverters) and infusion of diagnostic (such as contrast media) or therapeutic agents into the neuro, peripheral, and coronary vasculature. The catheter is inserted into the body through a small puncture made in the skin and into a blood vessel. The VIA 27 Microcatheter is contraindicated for use with liquid embolic materials, such as n-butyl 2-cyanoacrylate or ethylene vinyl alcohol & DMSO (dimethyl sulfoxide).

Why the product was recalled

Sequent Medical is recalling the VIA 27 Microcatheter because it may not adequately re-sheath the interventional device into the catheter as designed.

Distribution information

Worldwide Distribution - US including NY, MN, FL, MO, MD, TN, NC, OH, CA, KS, TN, WV and Internationally to Czech Republic, Germany, UK, France, Netherlands, Italy, Spain, Finland, Switzerland, Slovenia, Hungary, Northern Ireland, Turkey, and Denmark,

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.