FDA recall record

Zimmer VERSA-FX FEMORAL FIXATION SYSTEM COMPRESSION LAG SCREW, several length recall — FDA Z-2092-2013

Single packaging configuration used for the VERSA-FX-II Femoral Fixation System failed the leak test compromising the sterility of the product.

Class IIclassification
Z-2092-2013official record ID
Sep 4, 2013report date

Recall record at a glance

Official recordZ-2092-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZimmer, Inc.
Report dateSep 4, 2013
Recall initiation dateJun 19, 2013
Quantity85, 794 units in total
LocationWarsaw, IN, United States

Official product description

VERSA-FX FEMORAL FIXATION SYSTEM COMPRESSION LAG SCREW, several length (55 mm to 155 mm), diameters (12.7 mm and 15.8 mm) and short thread, standard thread and large thread. Item codes starting with 00-1193-005, 00-1193-010 and 00-1193-015. orthopedic implant.

Why the product was recalled

Single packaging configuration used for the VERSA-FX-II Femoral Fixation System failed the leak test compromising the sterility of the product.

Distribution information

Worldwide distribution: US Nationwide (including PR and USVI), and countries of: Israel, Argentina, Brazil, France, Mexico, Saudi Arabia, Beirut, China, Korea, San Salvador, Honduras, Canada, Australia, Switzerland, Germany, Arab Emirates, UK, Belgium, Egypt, Spain, France, Italy, Iran, Lebanon, Jordan, Iraq, Canary Islands.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0972-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0970-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0975-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0968-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.