FDA recall record

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA VERO Linear Accelerator System, Model Number recall — FDA Z-2653-2014

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of…

Class IIclassification
Z-2653-2014official record ID
Sep 24, 2014report date

Recall record at a glance

Official recordZ-2653-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Report dateSep 24, 2014
Recall initiation dateJul 5, 2014
Quantity3 units in use in the US
LocationHIROSHIMA, N/A, Japan

Official product description

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Why the product was recalled

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

Distribution information

Distributed in the states of NY, FL, and TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.