FDA recall record

Johnson & Johnson Surgical Vision VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS recall — FDA Z-2247-2023

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to…

Class IIclassification
Z-2247-2023official record ID
Aug 9, 2023report date

Recall record at a glance

Official recordZ-2247-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmJohnson & Johnson Surgical Vision, Inc.
Report dateAug 9, 2023
Recall initiation dateMay 30, 2023
Quantity148,941
LocationIrvine, CA, United States

Official product description

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Why the product was recalled

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Distribution information

Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Korea¿(South), Kuwait, Lebanon, Malaysia, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.