FDA recall record

Luminex Verigene Enteric Pathogens Nucleic Acid Test (EP), REF 30-002-023 recall — FDA Z-2023-2023

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test…

Class IIclassification
Z-2023-2023official record ID
Jul 5, 2023report date

Recall record at a glance

Official recordZ-2023-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmLuminex Corporation
Report dateJul 5, 2023
Recall initiation dateJun 5, 2023
Quantity3987 units
LocationNorthbrook, IL, United States

Official product description

Verigene Enteric Pathogens Nucleic Acid Test (EP), REF 30-002-023

Why the product was recalled

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.

Distribution information

Worldwide and US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 24, 2022

Luminex ARIES SARS-CoV-2 Assay recall — FDA Z-1524-2022

SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not…

Class IIFDA: OngoingLuminex Corporation