FDA recall record

Steris VERIFY Dual Species Self-Contained Biological Indicator recall — FDA Z-0072-2027

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…

Class IIclassification
Z-0072-2027official record ID
Oct 14, 2026report date

Recall record at a glance

Official recordZ-0072-2027
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSteris Corporation
Report dateOct 14, 2026
Recall initiation dateAug 31, 2026
Quantity1185
LocationMentor, OH, United States

Official product description

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

Why the product was recalled

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Distribution information

Worldwide - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records