Recall record at a glance
| Official record | Z-1929-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Philips Medical Systems (Cleveland) Inc. |
| Report date | Jul 10, 2019 |
| Recall initiation date | Jun 3, 2019 |
| Quantity | 70 |
| Location | Cleveland, OH, United States |
Official product description
Vereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecting, localizing, diagnosing, staging, re-staging, and follow-up for monitoring therapy response of various diseases in oncology, cardiology, and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.
Why the product was recalled
The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.
Distribution information
Worldwide Distribution - US Nationwide Domestic distribution to CA GA LA MD MI MN MO NE NJ NY OH TN TX VT WI. Foreign distribution to Mexico, Argentina Austria Belgium China Colombia France Germany Israel Italy Japan Latvia Luxembourg Netherlands Portugal Singapore Spain Switzerland
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.