FDA recall record

Olympus Corporation of the Americas Veran : Endobronchial Models: INS-5500 View Optical recall — FDA Z-2662-2023

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated…

Class IIclassification
Z-2662-2023official record ID
Oct 11, 2023report date

Recall record at a glance

Official recordZ-2662-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateOct 11, 2023
Recall initiation dateAug 10, 2023
Quantity6318 units
LocationCenter Valley, PA, United States

Official product description

Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC INS-5411 0.016" Stylet, 1235 mm L, 10/Bx INS-5410 SPiN Flex ANSO Needle, 22ga, 1.8mm OD, 5/Bx INS-5323 Sliding stopper INS-5300 "Always-On Tip Tracked Triple Needle Brush, 12mm L, 1.8mm OD, 5/Bx"

Why the product was recalled

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Distribution information

Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.