Recall record at a glance
| Official record | Z-1555-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | GE Medical Systems, LLC |
| Report date | Apr 23, 2025 |
| Recall initiation date | Mar 6, 2025 |
| Quantity | 12,808 total devices |
| Location | Waukesha, WI, United States |
Official product description
Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for China (Model Number H45113VGSU) diagnostic ultrasound systems
Why the product was recalled
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Distribution information
Worldwide - US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.