FDA recall record

US Endoscopy Group Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable recall — FDA Z-1943-2015

The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation…

Class IIclassification
Z-1943-2015official record ID
Jul 8, 2015report date

Recall record at a glance

Official recordZ-1943-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmUS Endoscopy Group Inc.
Report dateJul 8, 2015
Recall initiation dateFeb 17, 2015
Quantity2,173 units
LocationMentor, OH, United States

Official product description

Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.

Why the product was recalled

The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation of the components during use.

Distribution information

Worldwide Distribution - US Nationwide in the states of: CA, CO, IA, MA, OH, TX & WV and the countries of: Brazil, Costa Rica, Croatia, Czech Republic, Egypt, France, Germany, India, Indonesia, Italy, Latvia, Lithuania, Philippines, Poland, Romania, Russian Federation, Slovenia & South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records