FDA recall record

Navilyst Medical, Inc., an AngioDyamics Company Vaxcel PICC with PASV MST-30 Kit under the recall — FDA Z-1969-2016

PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis…

Class IIclassification
Z-1969-2016official record ID
Jun 22, 2016report date

Recall record at a glance

Official recordZ-1969-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNavilyst Medical, Inc., an AngioDyamics Company
Report dateJun 22, 2016
Recall initiation dateMar 14, 2016
Quantity1,786 units
LocationGlens Falls, NY, United States

Official product description

Vaxcel PICC with PASV MST-30 Kit under the following labels: 1) 4F, UPN M001454570, Rx ONLY, 2) 5F, UPN M001454620 & UPN M001454720, Rx ONLY & 3) 6F, UPN M001454770, Rx ONLY. The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intended for central venous access in adults, children and infants who require intravenous (IV) therapy.

Why the product was recalled

PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.

Distribution information

Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.