FDA recall record

Solta Medical VASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V recall — FDA Z-3048-2017

Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced…

Class IIclassification
Z-3048-2017official record ID
Aug 30, 2017report date

Recall record at a glance

Official recordZ-3048-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSolta Medical Inc.
Report dateAug 30, 2017
Recall initiation dateJun 1, 2017
Quantity7 units
LocationBothell, WA, United States

Official product description

VASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).

Why the product was recalled

Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.

Distribution information

US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.