Recall record at a glance
| Official record | Z-0054-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Terminated |
| Recalling firm | Vascular Solutions, Inc. |
| Report date | Oct 26, 2016 |
| Recall initiation date | Sep 16, 2016 |
| Quantity | 11,670 units total |
| Location | Maple Grove, MN, United States |
Official product description
Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature. The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment. The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters. Model 5200 also has hydrophilic coating on the distal 18cm of the catheter. The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively. The Twin-Pass catheter has a marker band located 1mm from the distal tip. Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.
Why the product was recalled
Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
Distribution information
Worldwide Distribution: US (nationwide) including Puerto Rico and outside the US to include: Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.