FDA recall record

Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology recall — FDA Z-0531-2013

Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality…

Class IIclassification
Z-0531-2013official record ID
Dec 26, 2012report date

Recall record at a glance

Official recordZ-0531-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVarian Medical Systems, Inc. Oncology Systems
Report dateDec 26, 2012
Recall initiation dateNov 5, 2012
Quantity264 units
LocationPalo Alto, CA, United States

Official product description

Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, Reference/FSCA Identifier: CP-09798; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA

Why the product was recalled

Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import.

Distribution information

Worldwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.