FDA recall record

Avid Medical Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory recall — FDA Z-1865-2017

Avid Medical Inc has received a complaint was received from an end-user indicating the vacutainer component 519700 (BD: 367863) expiring 12/16 was…

Class IIclassification
Z-1865-2017official record ID
Apr 26, 2017report date

Recall record at a glance

Official recordZ-1865-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAvid Medical, Inc.
Report dateApr 26, 2017
Recall initiation dateMar 15, 2017
Quantity25308
LocationToano, VA, United States

Official product description

Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory hematology studies including but not limited to lead level testing, FEP level testing, and CBC testing.

Why the product was recalled

Avid Medical Inc has received a complaint was received from an end-user indicating the vacutainer component 519700 (BD: 367863) expiring 12/16 was discovered inside of a kit labelled with an expiration date after 12/16. Upon receipt of complaint, AVID Medical promptly initiated an internal investigation and root cause analysis. Investigation found mixed component lots were contained in product 519700 component case. This occurrence resulted in the incorrect lot number and expiration being attributed to all work orders produced using the affected component. This issue was immediately contained and corrected. A full inventory analysis was conducted reviewing all vacutainer component inventories. The issue was found to affect a total of five vacutainer components.

Distribution information

All products are distributed through distribution centers confined to the US. All end-users for these products are also solely located in the US.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.