FDA recall record

Steris V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable recall — FDA Z-0737-2024

Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage…

Class IIclassification
Z-0737-2024official record ID
Jan 24, 2024report date

Recall record at a glance

Official recordZ-0737-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSteris Corporation
Report dateJan 24, 2024
Recall initiation dateNov 16, 2023
Quantity4,214 units
LocationMentor, OH, United States

Official product description

V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001

Why the product was recalled

Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 14, 2024

Steris Hookup recall — FDA Z-2538-2024

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an…

Class IIFDA: OngoingSteris Corporation