FDA recall record

Olympus Corporation of the Americas URF-P6 Uretero-reno fiberscope and URF-P6R Uretero-reno recall — FDA Z-1251-2017

Olympus is recalling the URF-P6/P6R endoscopes due to a similar device (URF-V2/V2R endoscopes) but not identical, where adverse event complaints…

Class IIclassification
Z-1251-2017official record ID
Mar 8, 2017report date

Recall record at a glance

Official recordZ-1251-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOlympus Corporation of the Americas
Report dateMar 8, 2017
Recall initiation dateDec 12, 2016
Quantity3461 units distributed to US consignees
LocationCenter Valley, PA, United States

Official product description

URF-P6 Uretero-reno fiberscope and URF-P6R Uretero-reno fiberscope

Why the product was recalled

Olympus is recalling the URF-P6/P6R endoscopes due to a similar device (URF-V2/V2R endoscopes) but not identical, where adverse event complaints associated with tissue trauma, perforation and insertion tubes which were stuck inside of patients that had to be surgically removed. In an effort to mitigate a potential risk to patient health, Olympus is undertaking this action to notify users of these complaints and the need for careful inspection of the endoscope prior to use in accordance with the instructions provided int heir notification letter.

Distribution information

Distributed Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.