FDA recall record

Zimmer Surgical UNIVERSAL Modular Electric/Battery Handpiece for surgery The electric recall — FDA Z-0165-2015

Non-injury complaints from Europe identified a malfunction that made it possible for a handpiece to start by itself when the power source is connected.

Class IIclassification
Z-0165-2015official record ID
Nov 12, 2014report date

Recall record at a glance

Official recordZ-0165-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZimmer Surgical Inc.
Report dateNov 12, 2014
Recall initiation dateOct 6, 2014
Quantity828
LocationDover, OH, United States

Official product description

UNIVERSAL Modular Electric/Battery Handpiece for surgery The electric handpiece is packaged within a foam bag inside of a padded-foam corrugate box. An operating manual is packed with the handpiece between the top foam pad and box lid. A product label is applied to the lid of the box. Product Usage: Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.

Why the product was recalled

Non-injury complaints from Europe identified a malfunction that made it possible for a handpiece to start by itself when the power source is connected.

Distribution information

US Nationwide Distribution in the states of: AK, AL,AR, AZ,CA,CO,CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA,ME, MI, MN, MO, NC, NE, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.