Beckman Coulter UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465 recall — FDA Z-2157-2017

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that…

Class IIclassification
Z-2157-2017official record ID
May 31, 2017report date

Recall record at a glance

Official recordZ-2157-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateMay 31, 2017
Recall initiation dateApr 24, 2017
Quantity4880 units total (2119 units in US)
LocationBrea, CA, United States

Official product description

UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.

Why the product was recalled

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.

Distribution information

Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote d'lvoire, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan , Kuwait, Lebanon, Lithuania ,Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania , Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia , South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay and Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.