FDA recall record

Shanghai United Imaging Healthcare Co uCT 760 System, Computed Tomography X-ray System recall — FDA Z-2522-2020

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and…

Class IIclassification
Z-2522-2020official record ID
Jul 15, 2020report date

Recall record at a glance

Official recordZ-2522-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmShanghai United Imaging Healthcare Co., Ltd.
Report dateJul 15, 2020
Recall initiation dateMay 8, 2020
Quantity5 units
LocationShanghai, China

Official product description

uCT 760 System, Computed Tomography X-ray System

Why the product was recalled

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Distribution information

Distributed to two customers. Both customers located in Houston, Texas.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.