Integra LifeSciences TS Series Depth Gauge for Integra Ti6 Internal Fixation System recall — FDA Z-1965-2016

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length…

Class IIclassification
Z-1965-2016official record ID
Jun 22, 2016report date

Recall record at a glance

Official recordZ-1965-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegra LifeSciences Corp.
Report dateJun 22, 2016
Recall initiation dateMay 10, 2016
Quantity71 units
LocationPlainsboro, NJ, United States

Official product description

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

Why the product was recalled

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Distribution information

US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.