FDA recall record

Clark Laboratories Trinity biotech Captia (TM) Syphilis IgM Capture Test Kit, Product recall — FDA Z-0819-2015

Through in-house routine real time stability monitoring of Trinity's Captia Syphilis IgM Capture Test Kit, Lot 033, the firm identified a failure of the…

Class IIIclassification
Z-0819-2015official record ID
Dec 24, 2014report date

Recall record at a glance

Official recordZ-0819-2015
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmClark Laboratories, Inc. (dba, Trinity Biotech USA)
Report dateDec 24, 2014
Recall initiation dateNov 17, 2014
QuantityDomestic: 9 kits; Foreign: 26 kits
LocationJamestown, NY, United States

Official product description

Trinity biotech Captia (TM) Syphilis IgM Capture Test Kit, Product No./REF 800-935, Test Kit for 96 Determinations, IVD for In Vitro Diagnostic use; an antibody class capture enzyme immunoassay for the detection of IgM antibodies to T. pallidum as an aid to the identification of active syphilis in human serum.

Why the product was recalled

Through in-house routine real time stability monitoring of Trinity's Captia Syphilis IgM Capture Test Kit, Lot 033, the firm identified a failure of the kit to meet its labeled expiry dating of 6 months.

Distribution information

New York, Florida, Utah, and California plus foreign distribution to Spain and France

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.