FDA recall record

INARI MEDICAL Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ recall — FDA Z-2299-2020

The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the…

Class IIclassification
Z-2299-2020official record ID
Jun 17, 2020report date

Recall record at a glance

Official recordZ-2299-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmINARI MEDICAL
Report dateJun 17, 2020
Recall initiation dateMar 23, 2020
Quantity318
LocationIrvine, CA, United States

Official product description

Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079

Why the product was recalled

The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged as another device must be used or suboptimal thrombus removal if the problem is not detected.

Distribution information

U.S Distribution in the states of .: IA, GA, MO, MI, AZ, FL, NJ, MN, LA, PA, AL, OH, NY, WI, CA, TX, TN, NH, IL, NC, CT, CO, OK, OR, IN, MA, VA, KY, WA, MS, SC, UT, SD, ND, AR, ID, DE, including DC.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.