Recall record at a glance
| Official record | Z-1043-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Stryker Howmedica Osteonics Corp. |
| Report date | Feb 11, 2015 |
| Recall initiation date | Dec 15, 2014 |
| Quantity | seven units |
| Location | Mahwah, NJ, United States |
Official product description
Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system is a two piece ( a shell and a liner) design that is assembled during surgery. The shell is press-fit into the patients prepared acetabulum and a linger (polyethylene or alumina ceramic) is then locked into the shell. The femoral head component of the total hip system articulates with the acetabular liner.
Why the product was recalled
Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present.
Distribution information
Nationwide Distribution including NY, OH, PA, and TX.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.