FDA recall record

CooperSurgical TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage recall — FDA Z-2364-2019

The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is…

Class IIclassification
Z-2364-2019official record ID
Sep 4, 2019report date

Recall record at a glance

Official recordZ-2364-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCooperSurgical, Inc.
Report dateSep 4, 2019
Recall initiation dateJul 8, 2019
Quantity49813
LocationTrumbull, CT, United States

Official product description

TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.

Why the product was recalled

The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.

Distribution information

Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 30, 2025

CooperSurgical HSG Procedure Tray recall — FDA Z-2135-2025

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

Class IIFDA: OngoingCooperSurgical, Inc.
Medical DeviceDec 25, 2024

K-Systems G85 Mini Incubator recall — FDA Z-0688-2025

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm…

Class IIFDA: OngoingCooperSurgical, Inc.
Medical DeviceDec 25, 2024

K-Systems T47 Warming Plate recall — FDA Z-0689-2025

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm…

Class IIFDA: OngoingCooperSurgical, Inc.
Medical DeviceDec 25, 2024

K-Systems Multiple/unknown recall — FDA Z-0692-2025

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm…

Class IIFDA: OngoingCooperSurgical, Inc.