FDA recall record

Philips Ultrasound Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as recall — FDA Z-1843-2020

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a…

Class IIclassification
Z-1843-2020official record ID
May 13, 2020report date

Recall record at a glance

Official recordZ-1843-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips Ultrasound Inc.
Report dateMay 13, 2020
Recall initiation dateAug 30, 2019
QuantityN/A
LocationBothell, WA, United States

Official product description

Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Why the product was recalled

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Distribution information

Worldwide distribution - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.