FDA recall record

Hill-Rom TotalCare SpO2RT2 bed recall — FDA Z-0949-2013

In certain situations a software problem with the TotalCare SpO2RT 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion…

Class IIclassification
Z-0949-2013official record ID
Mar 20, 2013report date

Recall record at a glance

Official recordZ-0949-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHill-Rom, Inc.
Report dateMar 20, 2013
Recall initiation dateFeb 26, 2013
Quantity309
LocationBatesville, IN, United States

Official product description

TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare SpO2RT 2 Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.

Why the product was recalled

In certain situations a software problem with the TotalCare SpO2RT 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the likelihood of skin irritation for patients at higher risk of developing pressure-related skin conditions. A correction to the software program is currently being developed and tested,

Distribution information

USA Nationwide Distribution including the state of Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 6, 2021

Hill-Rom LikoStretch Mod 600 IC recall — FDA Z-2601-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.
Medical DeviceOct 6, 2021

Hill-Rom LikoStretch 1900 recall — FDA Z-2600-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.