Recall record at a glance
| Official record | Z-1910-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Stryker Orthopaedics |
| Report date | Jun 5, 2024 |
| Recall initiation date | Apr 15, 2024 |
| Quantity | 334 units |
| Location | Montclair, NJ, United States |
Official product description
Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1. Software: TKA 2.0 + TKA 1.0.1 + THA 4.0.0.1 + PKA 3.0.2 Application Part Number: 700001590415
Why the product was recalled
Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment
Distribution information
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, Thailand, Turkey, UAE, and UK
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.