FDA recall record

Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for recall — FDA Z-2968-2020

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up…

Class IIclassification
Z-2968-2020official record ID
Sep 23, 2020report date

Recall record at a glance

Official recordZ-2968-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTosoh Bioscience Inc.
Report dateSep 23, 2020
Recall initiation dateAug 7, 2020
Quantity39 boxes
LocationGrove City, OH, United States

Official product description

Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 020661 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay

Why the product was recalled

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.

Distribution information

Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 2, 2019

Tosoh Bioscience ST AIA-PACK ACTH recall — FDA Z-0676-2019

ST AIA-PACK ACTH (ACTH: Adrenocorticotropic Hormone) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Adrenocorticotropic hormone (ACTH) in human EDT A…

Class IIFDA: TerminatedTosoh Bioscience Inc.