FDA recall record

Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in recall — FDA Z-3111-2017

It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital…

Class IIclassification
Z-3111-2017official record ID
Sep 20, 2017report date

Recall record at a glance

Official recordZ-3111-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmToshiba American Medical Systems
Report dateSep 20, 2017
Recall initiation dateMar 23, 2017
Quantity140
LocationTustin, CA, United States

Official product description

Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.

Why the product was recalled

It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error (2063)" which required a reboot and loss of the image.

Distribution information

USA (nationwide) Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.