Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000 is intended for use with recall — FDA Z-1175-2016

When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed…

Class IIclassification
Z-1175-2016official record ID
Mar 30, 2016report date

Recall record at a glance

Official recordZ-1175-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmToshiba American Medical Systems Inc.
Report dateMar 30, 2016
Recall initiation dateJan 14, 2016
Quantity28
LocationTustin, CA, United States

Official product description

Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000 is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.

Why the product was recalled

When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image. It also indicated to select the "OK" button to re-acquire the image data or to select the¿'Cancel" button to cancel the re-acquisition. As instructed the user selected "OK" and the same message window appeared. This same operation was repeated several times with the identical result - no image. Finally, the user selected "Cancel" and the re-acquisition was terminated.

Distribution information

Nationwide distribution to PA, UT, FL, NY, LA, WV, NJ, OH, WI and TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.