FDA recall record

Life Technologies Holdings Pte Torrent Suite Dx Software Version (IUO) 5.4 recall — FDA Z-0186-2020

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical…

Class IIclassification
Z-0186-2020official record ID
Oct 30, 2019report date

Recall record at a glance

Official recordZ-0186-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLife Technologies Holdings Pte Ltd
Report dateOct 30, 2019
Recall initiation dateSep 26, 2019
Quantitytotal of Ion+vela +chassis =400 units
LocationSingapore, Singapore

Official product description

Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15

Why the product was recalled

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Distribution information

states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.